Precision Trading PM5200 User Manual

Page 5

Advertising
background image

4

Blender

Air-Oxygen

WARNiNg

An Oxygen Analyzer/monitor must be used to verify

oxygen concentration.
Accuracy of oxygen concentration will be affected if bleed

is not activated at flow settings below 15 lpm for the High

Flow Blender, and 3 lpm for the Low Flow Blender.
dO NOT obstruct the alarm.
dO NOT use Blender when alarm is sounding.
dO NOT use oil in or around the Blender.
dO NOT occlude or obstruct the bleed port on the

auxiliary outlet of the Blender.
dO NOT use near any type of flame or flammable/

explosive substances, vapors or atmosphere.
Oxygen Concentration dial does not rotate 360 degrees.

Rotating the dial less than 21% or over 100% oxygen will

damage the Blender.




CAUTiON

Turn off gas supplies when Air-Oxygen Blender is not in use.
Store the Air-Oxygen Blender in a clean, dry area when

not in use.
The Air-Oxygen Blender contains magnetic, ferrous

material that may affect the results of an mRI.
Ensure all connections are tight and leak free.
Avoid excessive pressure surges greater than 100 psi

(6.9 bar) when pressuring the Blender inlets.
dO NOT steam autoclave.
dO NOT immerse Air-Oxygen Blender into any liquid.
dO NOT gas sterilize with (ETO) Ethylene Trioxide.
dO NOT use if dirt or contaminants are present on or

around the Blender or connecting devices.
dO NOT smoke in an area where oxygen is being

administered.
dO NOT clean with aromatic hydrocarbons.