Performance check, Cleaning – Precision Trading PM5200 User Manual

Page 12

Advertising
background image

11

Blender

Air-Oxygen

3. To activate the bleed, turn the knurled collar until the bleed

pin is engaged. Then slide the collar back until it contacts the

cover.

4. Confirm the concentration of air/oxygen with an Oxygen

Analyzer/monitor.

5. Confirm the flow of air and/or oxygen mixture to the patient.
6. Turn “OFF” the Air and Oxygen supply when deactivating the

Auxiliary bleed, and no device is attached to the outlet port.

7. To deactivate the bleed, pull the collar out away from the cover

while rotating the collar until the slot engages the pin. Then

slide the collar past the pin and turn an additional

1

/

4

turn to

lock in place.
NOTE: This Procedure is completed with ease when all gas

supplies to the Blender are turned “OFF”.

8. Turn “OFF” the Air and Oxygen supply or disconnect when

the Blender is not in use.

9. Operational Verification listed below should be done before

the Blender is placed in service.

Verification consists of:

• Alarm Test
• Verify FiO

2

Concentration with

a Oxygen Analyzer/monitor

• Reverse Gas Flow Procedure

PERFORMANCE CHECk

A detailed description of the tests listed below can be found in the

Blender Service manual (P/N 504827), available on the Internet;

www.precisionmedical.com

• Reverse Gas Flow Procedure
• Operational Verification Procedure (OVP)

CLEANiNg

CAUTiON

dO NOT steam autoclave.
dO NOT immerse the Air-Oxygen Blender into any liquid.
dO NOT use any strong solvent or abrasive cleaners.
dO NOT gas sterilize with (ETO) Ethylene Trioxide.
dO NOT clean with aromatic hydrocarbons.