Figure 2-11: setup loop–log values – RayTek 3i User Manual

Page 33

Advertising
background image

Raynger 3i Series Operator’s Manual

2-21

7. Set the TAM value by pressing the ▲ and ▼ buttons.

8. Press the ACTIVATE button to turn on the ambient temperature compensation

function. The Tamb icon will be displayed indicating that it is active. (Pressing
ACTIVATE again turns off the function.)

9. Pull trigger and reset the emissivity to the proper value for the target. You can

now take the target’s temperature using normal procedures. The current tempera-
ture and all computed values, as well as the analog and digital outputs, will now
be based on the compensated measurement.

Note: TAM only affects readings in the RUN loop; LOG loop readings are not affected.

2.4.7.2 SETUP Values for LOG

Figure 2-11 illustrates SETUP for the LOG loop.

To setup values for the LOG modes and functions, do the following:

1. Release or unlock the trigger, if necessary.

2. Press the SET button. The SET icon will be activated.

3. Press the RUN/LOG button, if necessary, so that the LOG icon is activated.

4. Press the ▲ or ▼ buttons to select the LOG location number (LOC mode).

MODE

RUN

LOG

SET

MODE

RECALL

ACTIVATE

LOCK

LASER

LOG

RCL

SET

°C

°F

HI ( ) LO ( ) Tamb

Press the

and

buttons to change the
Log Location (LOC)
number.

You can select and
log data in any location
number from 1 to 100.

Press the Set button
again, if the display
goes out, to continue
the SETUP functions

RUN

LOG

SET

MODE

RECALL

ACTIVATE

LOCK

LASER

LOG

RCL

SET

°C

°F

HI ( )

LO ( ) Tamb

Turn High Alarm
on or off with the
Activate button

Change the High Alarm
temperature value with
the

and

buttons

LOC Mode

HAL Mode

TRIGGER

Released & Unlocked

BUTTON

SET & LOG

LOOP ICON

LOG

SET

RUN

LOG

RECALL

MODE

ACTIVATE

LOCK

LASER

RUN

LOG

RECALL

MODE

ACTIVATE

LOCK

LASER

Figure 2-11: SETUP Loop–LOG Values

Advertising