LEAK TESTING

This function enables testing for leaks in pressure systems. The

instrument has a function to automate leak testing in a variety of

applications.

Automatic leak test results are logged, not stored and cannot be

output directly via HHP-2000DL into your MLWIN software. To send

leak test results to your PC please use your PC’s Hyper Terminal and

output as to an Epson Printer. When entering pressure values (P1,

P2, P3) you must enter numbers to exactly the same number of

decimal places as full pressure range of the instrument being used.

The number of decimal places required can be seen in the table of

pressure ranges on the back page.

LEAK TESTING

This function enables testing for leaks in pressure systems. The

instrument has a function to automate leak testing in a variety of

applications.

Automatic leak test results are logged, not stored and cannot be

output directly via HHP-2000DL into your MLWIN software. To send

leak test results to your PC please use your PC’s Hyper Terminal and

output as to an Epson Printer. When entering pressure values (P1,

P2, P3) you must enter numbers to exactly the same number of

decimal places as full pressure range of the instrument being used.

The number of decimal places required can be seen in the table of

pressure ranges on the back page.

AVERAGING

MAX/MIN

MAX/MIN

Press

key (when in

mode) and

symbol, together

with average reading, will appear on display. Press

to return to

mode.

AVERAGING

MAX/MIN

MAX/MIN

Press

key (when in

mode) and

symbol, together

with average reading, will appear on display. Press

to return to

mode.

SETTING UP

Press

key (when not in

mode) and

and

symbols will appear on the display. Enter pre-stress pressure level

(if pre-stress pressure function is not required, press

key) and the

display will show

and

symbols. Key in start pressure and

press

key and the

and

symbols will now be displayed.

Key in test time in minutes and seconds (MM.SS), press

key and

the

and

symbols will now be displayed. Key in permissable

change in pressure and press

key. The instrument is now ready

to run test and

will appear on display.

MAX/MIN

NE

NE

P1

P2

NE

NE

P3

P1

SETTING UP

Press

key (when not in

mode) and

and

symbols will appear on the display. Enter pre-stress pressure level

(if pre-stress pressure function is not required, press

key) and the

display will show

and

symbols. Key in start pressure and

press

key and the

and

symbols will now be displayed.

Key in test time in minutes and seconds (MM.SS), press

key and

the

and

symbols will now be displayed. Key in permissable

change in pressure and press

key. The instrument is now ready

to run test and

will appear on display.

MAX/MIN

NE

NE

P1

P2

NE

NE

P3

P1

X

X

X

X

X

X

OUTPUT OF DATA

1

2

3

This feature will only function if either the

or

symbols are

displayed.

To view data press

key, then and keys. The first stored

reading will be displayed. Press

key to view successively stored

readings and any other key to return to normal mode.

To output data to a PC press

key, then key.

Ensure PC/Printer is ready to

receive information.

After transmission, the unit will return to normal operation and any

auto logging will stop.

To erase stored data press

key, then and

keys.

There will be a delay before the unit will accept the

key and

symbol

will be displayed. This is to help stop accidental erasure.

After erasure, neither

or

symbols will be displayed. The unit

will return to normal mode.

The selected transmit option will be displayed during transmission.

If after pressing key you do not wish to

delete data, you must turn instrument off and on, then the stored

data will not be lost.

To output data to an Epson compatible printer press

key, followed

by key. Printer settings must be as follows: 9600 Baud, 8 data bits,

No Parity, One stop bit, No flow control. It is also possible to print to

Hyper Terminal in your PC in this way.

4

3

OUTPUT OF DATA

1

2

3

This feature will only function if either the

or

symbols are

displayed.

To view data press

key, then and keys. The first stored

reading will be displayed. Press

key to view successively stored

readings and any other key to return to normal mode.

To output data to a PC press

key, then key.

Ensure PC/Printer is ready to

receive information.

After transmission, the unit will return to normal operation and any

auto logging will stop.

To erase stored data press

key, then and

keys.

There will be a delay before the unit will accept the

key and

symbol

will be displayed. This is to help stop accidental erasure.

After erasure, neither

or

symbols will be displayed. The unit

will return to normal mode.

The selected transmit option will be displayed during transmission.

If after pressing key you do not wish to

delete data, you must turn instrument off and on, then the stored

data will not be lost.

To output data to an Epson compatible printer press

key, followed

by key. Printer settings must be as follows: 9600 Baud, 8 data bits,

No Parity, One stop bit, No flow control. It is also possible to print to

Hyper Terminal in your PC in this way.

4

3

Press

key.

symbol and

will appear on the display.

Enter required time interval in hh:mm (ie, in order to take a reading

every 1 minute, key in 0001) and press

key. Current reading will

return to display and

symbol will remain on display.

NE

0000

Instrument

will continue logging until a maximum of 250 readings are taken or

data is down-loaded.

Press

key.

symbol and

will appear on the display.

Enter required time interval in hh:mm (ie, in order to take a reading

every 1 minute, key in 0001) and press

key. Current reading will

return to display and

symbol will remain on display.

NE

0000

Instrument

will continue logging until a maximum of 250 readings are taken or

data is down-loaded.

WARRANTY / DISCLAIMER

RETURN REQUESTS / INQUIRIES

OMEGA ENGINEERING, INC warrants this unit to be free of defects in materials and

workmanship for a period of 25 months from the date of purchase. OMEGA Warranty

adds an additional one (1) month grace period to the normal one (1) year product

warranty to cover handling and shipping time. This ensures that OMEGA’S

customers receive maximum coverage on each product. If the unit malfunctions, it

must be returned to the factory for evaluation. OMEGA’s Customer Service Depart-

ment will issue an Authorized Return (AR) number immediately upon phone or written

request. Upon examination by OMEGA, if the unit is found to be defective, it will be

repaired or replaced at no charge. OMEGA’S WARRANTY does not apply to defects

resulting from any action of the purchaser, including but not limited to mishandling,

improper interfacing, operation outside of design limits, improper repair, or

unauthorized modification. This WARRANTY is VOID if the unit shows evidence of

having been tampered with or shows evidence of having been damaged as a result of

excessive corrosion; or current, heat, moisture or vibration; improper specification;

misapplication; misuse or other operating conditions outside of OMEGA’s control.

Components which wear are not warranted, including but not limited to contact points,

fuses, and triacs. OMEGA is pleased to offer suggestions on the use of its various

products. However, OMEGA neither assumes responsibility for any omissions or

errors nor assumes liability for any damages that result from the use if its products in

accordance with information provided by OMEGA, either verbal or written. OMEGA

warrants only that the parts manufactured by it will be as specified and free of

defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF

ANY KIND WHATSOEVER, EXPRESS OR IMPLIED, EXCEPT THAT OF TITLE,

AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANT-

ABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DIS-

CLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth herein

are exclusive, and the total liability of OMEGA with respect to this order, whether

based on contract, warranty, negligence, indemnification, strict liability or otherwise,

shall not exceed the purchase price of the component upon which liability is based.

In no event shall OMEGA be liable for consequential, incidental or special damages.

CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be

used: (1) as a “Basic Component” under 10 CFR 21 (NRC), used in or with any

nuclear installation or activity; or (2) in medical applications or used on humans.

Should any Product(s) be used in or with any nuclear installation or activity, medical

application, used on humans, or misused in any way, OMEGA assumes no

responsibility as set forth in our basic WARRANTY / DISCLAIMER language, and,

additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from any

liability or damage whatsoever arising out of the use of the Product(s) in such a

manner.

Direct all warranty and repair requests/inquiries to the OMEGA Customer Service

Department. BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER

MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S

CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING

DELAYS). The assigned AR number should then be marked on the outside of the

return package an on any correspondence.

FOR WARRANTY RETURNS,

For NON-WARRANTY REPAIRS,

please have the following information

consult OMEGA for current repair

available BEFORE contacting OMEGA:

charges. Have the following information

1. Purchase Order number under which

available BEFORE contacting OMEGA:

the product was PURCHASED,

1. Purchase Order number to cover the

2. Model and serial number of the

COST of the repair/calibration

product under warranty, and

2. Model and serial number of the

3. Repair instructions and/or specific

product, and

problems relative to

3. Repair instructions and/or specific

problems relative to the product.

OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible.

This affords our customers the latest in technology and engineering.

OMEGA is a registered trademark of OMEGA ENGINEERING, INC.

© Copyright 1998 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied,

photocopied, reproduced, translated, or reduced to any electronic medium or machine-readable form, in

whole or in part, without the prior written consent of OMEGA ENGINEERING, INC. It is the policy of

OMEGA to comply with all worldwide safety and EMC / EMI regulations that apply. OMEGA is constantly

pursuing certification of its products to the European New Approach Directive. OMEGA will add the CE

mark to every appropriate device upon certification. The information contained in this document is believed

to be correct, but OMEGA Engineering, Inc. accepts no liability for any errors it contains, and reserves the

right to alter the specifications without notice. WARNING: These products are not designed for use in, and

should not be used for, patient connected applications.

WARRANTY / DISCLAIMER

RETURN REQUESTS / INQUIRIES

OMEGA ENGINEERING, INC warrants this unit to be free of defects in materials and

workmanship for a period of 25 months from the date of purchase. OMEGA Warranty

adds an additional one (1) month grace period to the normal one (1) year product

warranty to cover handling and shipping time. This ensures that OMEGA’S

customers receive maximum coverage on each product. If the unit malfunctions, it

must be returned to the factory for evaluation. OMEGA’s Customer Service Depart-

ment will issue an Authorized Return (AR) number immediately upon phone or written

request. Upon examination by OMEGA, if the unit is found to be defective, it will be

repaired or replaced at no charge. OMEGA’S WARRANTY does not apply to defects

resulting from any action of the purchaser, including but not limited to mishandling,

improper interfacing, operation outside of design limits, improper repair, or

unauthorized modification. This WARRANTY is VOID if the unit shows evidence of

having been tampered with or shows evidence of having been damaged as a result of

excessive corrosion; or current, heat, moisture or vibration; improper specification;

misapplication; misuse or other operating conditions outside of OMEGA’s control.

Components which wear are not warranted, including but not limited to contact points,

fuses, and triacs. OMEGA is pleased to offer suggestions on the use of its various

products. However, OMEGA neither assumes responsibility for any omissions or

errors nor assumes liability for any damages that result from the use if its products in

accordance with information provided by OMEGA, either verbal or written. OMEGA

warrants only that the parts manufactured by it will be as specified and free of

defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF

ANY KIND WHATSOEVER, EXPRESS OR IMPLIED, EXCEPT THAT OF TITLE,

AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANT-

ABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DIS-

CLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth herein

are exclusive, and the total liability of OMEGA with respect to this order, whether

based on contract, warranty, negligence, indemnification, strict liability or otherwise,

shall not exceed the purchase price of the component upon which liability is based.

In no event shall OMEGA be liable for consequential, incidental or special damages.

CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be

used: (1) as a “Basic Component” under 10 CFR 21 (NRC), used in or with any

nuclear installation or activity; or (2) in medical applications or used on humans.

Should any Product(s) be used in or with any nuclear installation or activity, medical

application, used on humans, or misused in any way, OMEGA assumes no

responsibility as set forth in our basic WARRANTY / DISCLAIMER language, and,

additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from any

liability or damage whatsoever arising out of the use of the Product(s) in such a

manner.

Direct all warranty and repair requests/inquiries to the OMEGA Customer Service

Department. BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER

MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S

CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING

DELAYS). The assigned AR number should then be marked on the outside of the

return package an on any correspondence.

FOR WARRANTY RETURNS,

For NON-WARRANTY REPAIRS,

please have the following information

consult OMEGA for current repair

available BEFORE contacting OMEGA:

charges. Have the following information

1. Purchase Order number under which

available BEFORE contacting OMEGA:

the product was PURCHASED,

1. Purchase Order number to cover the

2. Model and serial number of the

COST of the repair/calibration

product under warranty, and

2. Model and serial number of the

3. Repair instructions and/or specific

product, and

problems relative to

3. Repair instructions and/or specific

problems relative to the product.

OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible.

This affords our customers the latest in technology and engineering.

OMEGA is a registered trademark of OMEGA ENGINEERING, INC.

© Copyright 1998 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied,

photocopied, reproduced, translated, or reduced to any electronic medium or machine-readable form, in

whole or in part, without the prior written consent of OMEGA ENGINEERING, INC. It is the policy of

OMEGA to comply with all worldwide safety and EMC / EMI regulations that apply. OMEGA is constantly

pursuing certification of its products to the European New Approach Directive. OMEGA will add the CE

mark to every appropriate device upon certification. The information contained in this document is believed

to be correct, but OMEGA Engineering, Inc. accepts no liability for any errors it contains, and reserves the

right to alter the specifications without notice. WARNING: These products are not designed for use in, and

should not be used for, patient connected applications.

Advertising