SPECIFICATIONS

Typical @ 25 C and rated supply voltage unless otherwise specified.


INPUTS:

• Input types: J,K,T,E,R,S,B t/c, RTD, Voltage, Millivolt, Current & thermistor

• Cold junction compensation error: +/- 0.5C max (10C to 40C)
• Open thermocouple indication: ‘OPEN TC’ displayed


OPTIONS:

• 240Vac @ 50 Hz Power Supply

• 15Vdc @ 1000ma. Power Supply

• Open Collector output: 6 open collector transistors @ 50ma. sink each
• Relay output: 6 SPST 1 Amp. @ 28Vdc, 0.5 Amp. @ 120Vac (resistive load)

(for 220 Vac units only 0.25Amp.@ 240Vac (resistive load))

ACCURACY:

• Temperature resolution: 1 C/1 F, 0.1C for Thermistor

• 0.1% of Full Scale (Base metal T/C), 0.5% others

ANALOG TO DIGITAL CONVERSION:

• 4 1/2 Digit (20000 count) A/D conversion.

• Dual slope integrating converter.

• Conversion Rate: 7 /sec. typical

POWER REQUIREMENTS:

• 120Vac, 0.2 Amp. @ 60 Hz


DIMENSIONS:

Case: 7.55"W x 6.29"H x 4.05"D

IP65 rated plastic enclosure

• Weight: 2.63 lb. (1.19kg.)

• Enclosure: All Plastic enclosure—IP65 Ingress Protection

DISPLAY:

• Red 7-segment LED display, 0.39 inch (10mm) digit height

• Negative polarity indication
• Out of range indication: HELP

• Display Test: Briefly displays 8.8.8.8.8.8.8. on power up


RELIABILITY:

• Calibration: NIST traceable
• Recalibration: Recommended at 12 months interval