Specifications for gas units – Omega Engineering FLR1000 User Manual

Page 17

Advertising
background image

pg. 17 of 29

E.

Specifications For Gas Units


FLR1000

FLR1000-D

FLR1000-BR-D

Display None

3.5 Digit LCD,

0.39” (10mm) high digits

Accuracy
(including linearity)

±3.0% Full Scale*

Repeatability

±0.5% Full Scale*

Pressure Rating

40 psig

(2.8 bar)

Temperature Rating

Operating Range: 5 to 55ºC

Storage Range: 0 to 70ºC

Temperature
Sensitivity

±0.2% F.S.* or less per ºC

Wetted Materials

Ryton

®

316 Stainless*

Epoxy

Glass

Sapphire

Brass

Ryton

®

316 Stainless*

Epoxy

Glass

Sapphire

O-Ring Material

Standard - Viton

®

EPDM Optional

Fitting Material
(Standard)

Acetal Brass

Fitting Material
Optional

Acetal, brass, or stainless steel

Recommended
Filtration

25 microns or less

Compatible Gases

Clean, dry

0-5VDC Output Signal

Standard

Minimum 2.5 KΩ load

Typical Power
Consumption

Standard: 12 VDC @ 35 mA (11.5-15 VDC)

Model 110 Power Adapter 115/230VAC

Response Time

Typically <30 seconds for 67% of final value

Reliability 100,000

hours

MTBF

Certifications CE

Approved

89/336/EEC (EN 55011 & EN 50082-1)

73/23/EEC Low Voltage Directive

Ratings

IP10 (NEMA 1)

Warranty

1 year limited

*Specifications from 20-100% of rated flow. Linearity is best fit straight line. All calibrations
performed with AIR unless otherwise stated on calibration certificate

Advertising