Omega Engineering FLR1000 User Manual

Page 27

Advertising
background image

pg. 27 of 29

Symptom

Possible Cause

Method of Correction

Flow indication is not

linear and output too
low at lower flows

Attempting to measure

flows below the specified
minimum for the unit

Moisture or liquid in sensor



Mechanical sensor assembly
damaged

Use higher flow rates




Remove moisture and allow

sensor to dry out

Unit must be returned for
repair (see Section J)

Flow reading accuracy

is outside the
specification

Gas has a specific gravity

other than Air at 20°C

Moisture in sensor


Mechanical sensor assembly
damaged

Pressure drop issue

Calculate the flow reading

error / offset as explained in
section C6

Allow sensor to dry out.
Ensure gas is dry.

Unit must be returned for
repair (see Section J)

Correct for the pressure
drop ( ΔP )

Flow indication is fixed

at one value

The maximum rated flow for

the unit is being exceeded.

Output circuits may be
damaged

Reduce the flow


Unit must be returned for
repair (see Section J)