Troubleshooting guide – Omega Engineering FLR1000 User Manual

Page 26

Advertising
background image

pg. 26 of 29

H.

Troubleshooting Guide

Symptom

Possible Cause

Method of Correction

Unit Leaks

Fittings not tight enough.

Sensor assembly cracked

Tighten fittings (see section
B4)

Unit must be returned for
repair (see Section J)

No output signal or
flow indication

No power or low power

No flow passing through
unit



Output shorted or

overloaded


Improper connection




Moisture on internal

electronics


Output circuitry damaged


Apply correct power

Display or output should
read zero.



Check electrical connections

(see section B5)


Check electrical connections
(see section B5) and ensure

connector is correctly
plugged in.

Allow unit to dry out in a

dry environment.


Unit must be returned for
repair (see Section J)


Flow indication
“bounces” or
fluctuates

Flow is actually fluctuating

Using a pulsatile or piston
pump that causes a
pulsating flow of gas

Not using the fittings
supplied by the
manufacturer

Moisture in sensor



Turbulence in tubing or

sensor


Fluctuating input power



Mechanical sensor assembly
damaged

Correct flow source to
obtain steady flow

Switch to a stable source of
gas


Use the supplied fittings



Remove moisture and allow

unit to dry out

Straighten the inlet tubing

or lengthen it to at least 6”
(150mm)

Correct the power source or
change to a regulated

supply

Unit must be returned for
repair (see Section J)