Section 5 Options and Accessories

5-22

probe. Use only 2-mil membranes with D.O. probes
connected to Isco flow meters.

Figure 5-6 D.O. Parameter Module

You must use the Isco 270 D.O. Module box between the probe
and flow meter; this extends the distance between the probe and
the flow meter to 1,000 feet.

5.10.6 Calibrating the D.O.

Probe with a Flow
Meter

Prepare the probe as described above if this has not already been
done (fill the probe cavity with electrolyte and seat the mem-
brane).

Table 5-4 D.O. Probe Specifications

Cathode Gold

Anode

Silver

Membrane

FEP Teflon; 2 mil standard

Electrolyte

Half-saturated KCl (Potassium Chloride)

Temp. Range

0° to +45° C

Polarizing Voltage

0.8 Volts (Nominal)

Probe Current in Air at 30° C

19 microamps (nominal)

Probe Current in Nitrogen at 30° C

0.15 microamps or less

Connection

12 ft. (3.2 m) cable with 5-pin male M/S connec-
tor

Response Time

Typical response for dissolved oxygen, using
supplied membranes, is 90% in 20 seconds.
Response at low dissolved oxygen levels is typ-
ically 90% in 60 seconds.