4220 Flow Meter
Section 2 Programming

2-42

2.4.6 Step 4 - Reset Totalizer

This step allows you to reset the flow meter's internal flow
totalizer. Note that there is the possibility of more than one
totalizer. All models have the capability of maintaining a sep-
arate totalizer for the time the sampler is enabled through the
sampler enabling feature (step 6).

If you select step 4, the following will appear:

If you select NO, the flow meter will advance to the next step. If
you select YES, the flow meter will reset the internal totalizer to
zero.

Then, if you have sampler enabling turned on (step 6), the flow
meter will ask you whether you want to reset the totalizer for the
total flow during the time the sampler was enabled. (See step 6
for an explanation of sampler enabling.)

Then:

2.4.7 Step 5 - Sampler

Pacing

This step determines how the flow meter will signal an asso-
ciated automatic wastewater sampler to take a sample. The flow
meter and sampler must be connected together with a cable. The
flow meter sends flow pulses to the sampler. The sampler uses
these flow pulses as counts. When an appropriate number of flow
pulses has been received by the sampler, it will take a sample. If
you select step 5, the following will appear:

VOLUME will not appear if the flow meter is measuring level
only, or is only being used for parameter monitoring.

The Flowlink menu option will not appear unless Flowlink
software is installed and pacing has been downloaded from
Flowlink.

If you select DISABLE for sampler pacing, the flow meter will be
effectively disconnected from the sampler. The sampler will run
its program as if the flow meter weren't there. Selection of
DISABLE will also cause the flow meter to advance to the next
Program step. If you select VOLUME for sampler pacing, the fol-
lowing will appear:

• Other units of measure may appear here. The range for

pacing is max. flow divided by 100 (minimum) or 10,000
times max. flow (maximum.)

If Flowlink appears for sampler pacing, the operation of sampler
pacing has been determined by choices made in Flowlink, Isco's
proprietary data acquisition and storage software. Flowlink con-

RESET TOTALIZER
• YES • • NO •

ENABLE TOTALIZER

XX

CF (or other units of measure)

PRESS 'ENTER

RESET SAMPLER ENABLE TOTALIZER
• YES • • NO •

SAMPLER PACING
• DISABLE • • (VOLUME) • • (FLOWLINK) • • CONDITIONAL •

SAMPLER PACING
ENTER PACING VOLUME XX.XXXX CF