4220 Flow Meter
Section 1 Introduction

1-8

1.6 Technical

Specifications

The technical specifications for the 4220 Flow Meter are found in
Table 1-2. The anticipated longevity for a roll of paper used in the
internal plotter is shown for various chart speeds in Table 1-4.
Battery life expectancy is found in Table 1-5.

Table 1-2 Technical Specifications for the 4220 Flow Meter

Physical Specifications

Size 16"

high

× 11

1

2

"

wide

× 10

1

2

"

deep (40.6 Ч 29.2 Ч 26.7 cm) (without

power source attached).

Weight

17 lb. 4 oz. (7.7 kg)

Material

High-impact molded polystyrene structural foam.

Type

Self-certified NEMA 4X Enclosure.

Display Type

2-line, 40 character/line alphanumeric dot matrix liquid crystal.

Power

12 -14 VDC, 16 mA average at 12.5 VDC (Plotter set at 1" per hour).

Typical Battery Life

7-10 days with plotter at 1" per hour and 4 Ampere-hour Ni-Cad bat-
tery.

14 days with plotter turned off.

Operating Temperature

0° to 140°F (–18° to 60°C).

Storage Temperature

–40° to 158°F (–40° to 70°C).

Additional Power Required

for Optional Equipment

Modem

60 mA maximum during operation; 0.1 mA maximum standby.

High-Low Alarm Relay Box

10 mA standby, typical; 190 mA - both relays operated.

Internal Printer

Chart Speeds

Off,

1

2

, 1, 2, or 4 inches per hour.

Ribbon

19.7 ft. (6 m) black nylon - replaceable.

Operating Speed

1.5 lines per second at 68° F. (20°

C).

Character Size

0.09" high

× 0.07" wide (2.4mm × 1.7 mm), 12 pitch.

Printer Recording Span

Chart Resolution

Display Resolution

User-selected from

1

4

ft. (3.6 cm) to over 30 ft. (9.1 m) with multiple

over-ranges.

1

240

of selected recording span

0.001 ft. (0.3 mm)

Paper 4.5"

wide

× 65 ft. (11.4 cm × 19.8 m) plain white paper, replaceable.

Printer Reliability

2.5 million lines MCBF (mean cycles before failure).

Reports Printed

Program selections, interval activity reports, flow meter history.

Printer Recording Modes

Level, flow rate; includes totalized flow, sampler events, rainfall, tem-
perature, velocity, pH (or) D. O. (dissolved oxygen).

Plotted Linear Data

3 different linear plots can be printed at the same time.