Caution, 4 grounding – Toshiba GF632 User Manual

Page 26

Advertising
background image


- 25 -

6F8A0883

4.4 Grounding

CAUTION

„ Do not wire cables and replace parts

when power is supplied.

„ Do not work on piping and wiring with wet hands.

DON’T

Wiring work and replacing
parts in the power-on state
may cause electric shock.

DON’T

Wet hands may result in electric shock.

(1) Grounding of the GF630/LF600F and GF630/LF610F type

Ground as shown in Figure 4.12. Make the grounding wire as short as possible. Use grounding
wire material of IV wire 5.5mm

2

or more. Do not share a grounding wire with other

instruments where grounding current may flow. (An independent grounding is preferable.)





Figure 4.12 Grounding the GF630/LF600F and GF630/LF610F Type


Please connect the ground cables
of detector to piping using the M5 screw
shown as Figure4.13 as needed.
(In the case that there are no threaded hole

with piping, please tap the threaded hole.)

※It is necessary to connect the ground cable
surely electrically.

grounding resistance
100ohm or less

•In the case that the piping is conductive material,
connect the ground cables to the both ends of the
piping flange.

• In the case that the piping is non-conductive material,
use grounding rings.

Integral

Conductive material pipe
Example: Metal, etc.

Piping of non-conductive material
Example: Resin or metal piping whose
inside is resin-lined, etc.

Grounding wire

grounding ring

Grounding terminal

grounding resistance
100ohm or less

A

Figure4.13 Grounding wire connection

(Detail of part A)

M5

Ground cable of detector