Operating Basics

80A03 TekConnect Probe Interface Module Instruction Manual

15

Table 3: Recommended equipment for triggering setups (cont.)

To trigger only

To trigger and view a signal

Compatible TekConnect probe.
See page 1.

Compatible TekConnect probe.
See page 1.

(1) SMA cable, 0.5 m

1

(3) SMA cables, 0.5 m

1

Torque wrench, SMA 5/16, 8 in-lb

Torque wrench, SMA 5/16, 8 in-lb

------------------------------------------------------------

(1) 6 dB power divider

2

1

Tektronix part number 174-1427-00.

2

Tektronix part number 015-0565-00.

To set up the TekConnect Probe Interface Module to trigger CSA8000 and
TDS8000 Series Oscilloscopes, do the following:

1.

Remove the semi-rigid cable that connects the CH B TekConnect Probe
Interface Module to the sampling module. See Figure 8.

Figure 8: Removing the semi-rigid cable

2.

Use a 0.5 m SMA cable to connect the CH B TekConnect Probe Interface
Module to the sampling oscilloscope trigger input (Trigger Direct or Trigger
Prescaled). See Figure 9 on page 16.

3.

Attach the TekConnect measurement probe to CH A on the TekConnect
Probe Interface Module. See Figure 9.

4.

Attach a TekConnect trigger probe to CH B on the TekConnect Probe
Interface Module. See Figure 9.

Triggering CSA8000 and

TDS8000 Series

Oscilloscopes