Operating Basics

18

80A03 TekConnect Probe Interface Module Instruction Manual

To set up the TekConnect Probe Interface Module to trigger CSA8000 and
TDS8000 Series Oscilloscopes, do the following:

1.

Remove the semi-rigid cable that connects the CH B TekConnect Probe
Interface Module to the sampling module. See Figure 8 on page 15.

2.

Use a 0.5 m SMA cable to connect the 6 dB (2X) power divider to the
disconnected CH B SMA connector on the TekConnect Probe Interface
Module. See Figure 11.

3.

Use a 0.5 m SMA cable to connect the disconnected port on the sampling
module to one side of the 6 dB (2X) power divider. See Figure 11.

4.

Use a 0.5 m SMA cable to connect the other side of the power divider to the
sampling oscilloscope trigger input (Trigger Direct or Trigger Prescaled).

5.

Attach a TekConnect trigger probe to CH B on the TekConnect Probe
Interface Module. See Figure 11.

CSA8000/TDS8000

6 dB power

divider

Trigger direct input

80A03

Trigger prescaled

input

Circuit under test

Trigger and
Measurement probe

Figure 11: Setup for triggering the sampling oscilloscope and viewing signals using
the 80A03 TekConnect Probe Interface Module

Triggering and Viewing

Signals on CSA8000 and

TDS8000 Series

Oscilloscopes