Basic operation – Teledyne Trace Oxygen Analyzer 3000TA User Manual

Page 11

Advertising
background image

Model 3000TA

Quickstart Guide

Teledyne Analytical Instruments

11

BASIC OPERATION

Once the analyzer has been installed, it can be configured for your application. Prior to

using the instrument you must:

Calibrate the instrument

Define the analysis ranges and choose whether to use autoranging or a fixed range.

Set alarm setpoints and modes of alarm operation.

Additional configuration parameters are available to tailor the instrument to your particular

application. Refer to the Instruction Manual for specific information on setting of additional
system parameters.

Default Configuration

The default analyzer configuration is as follows:

Ranges: LO = 100 ppm, MED = 1000 ppm, HI = 10,000 ppm.

Auto Ranging: ON

Alarm Relays: Defeated, 1000 ppm, HI, Not failsafe, Not latching.

Zero: Auto, every 0 days at 0 hours.

Span: Auto, at 000,008.00 ppm, every 0 days at 0 hours.

If you choose not to use password protection, the default password is automatically

displayed on the password screen when you start up, and you simply press Enter for access to all
functions of the analyzer.

Using the Data Entry and Function Buttons

The ◄► arrow buttons select options from the menu currently being displayed on the

display. When selected, the option blinks.

When the selected option includes a modifiable item, the ▲▼ arrow buttons can be used to

increment or decrement that modifiable item.

The Enter button is used to accept any new entries on the VFD screen.
The Escape button is used to abort any new entries on the VFD screen that are not yet

accepted by use of the Enter button.