Quickstart Guide

Trace Oxygen Analyzer


Teledyne Analytical Instruments

8

Connect primary input power:
Insert the power cord into the power cord receptacle.

Install the fuse:
Fuses are not installed at the factory. Be
sure to install the proper fuse as part of
installation.

Place small screwdriver in notch, and pry
cover off, as shown.

To change between American and
European fuses, remove the single retaining
screw, flip fuse block over 180 degrees,
and replace screw.

Replace fuse as shown below and reassemble housing.

Connect Analog Outputs
There are four DC output signal connectors with spring terminals on the rear panel
where the analog output signals are available. There are two wires per output with the
polarity noted. See Section 3.3.2 in the Instruction Manual for detailed descriptions of
the analog outputs.

Connect Alarm Relays:
Use the three alarm-circuit connector spring terminals on the rear panel for making
connections to internal alarm relay contacts.

Connect Digital Remote Cal Inputs:
These input terminals accept 0 V (off) or 24 V dc (on) inputs for remote calibration
control.

Connect Cal Contact:
This relay contact is closed while analyzer is spanning and/or zeroing.