Model 3000TA

Quickstart Guide

Teledyne Analytical Instruments

7

4. If using remote calibration feature:

Connect zero gas to the ZERO IN port.

Connect span gas to the SPAN IN port.

5. Otherwise, calibration gases must be Tee'd into the sample inlet with appropriate

valves.

6. Regulate the sample gas pressure between 2 and 50 psig sufficient to keep the front

panel flowmeter reading in an acceptable range (0.1 to 2.4 SLPM). For non-
pressurized sample or very low pressure, (less than 2 psig) vacuum service plumbing
is recommended.

7. If greater sample flow is required for improved response time, install a bypass in the

sampling system upstream of the analyzer input.

8. For optional vacuum application where the sample pressure is at atmospheric or very

low pressure, the instrument should be fitted with the vacuum service option. This
option locates the flow control valve on the exhaust side of the Micro-fuel Cell.

Note: Exhaust connections must be consistent with the hazard level of the constituent

gases. Check Local, State, and Federal laws, and ensure that the exhaust
stream vents to an appropriately controlled area, if required.

Electrical Connections

CAUTION:

READ ALL SAFETY INFORMATION REGARDING ELECTRICAL HAZARDS
ASSOCIATIED WITH THIS INSTRUMENT BEFORE MAKING ELECTRICAL
CONNECTIONS.

CAUTION:

POWER IS APPLIED TO THE INSTRUMENT'S CIRCUITRY AS LONG AS THE
INSTRUMENT IS CONNECTED TO THE POWER SOURCE. THE RED
SWITCH ON THE FRONT PANEL IS FOR SWITCHING POWER ON OR OFF
TO THE DISPLAYS AND OUTPUTS ONLY.