Operation and Description of Instrument Function

VH87 Element Interval

Item type: read and write

Default: 1000mm

Range: 0mm ~ 99999mm

Applies only when Even Interval spacing is selected.

Caution!

Changing element interval, setting element position, these are only applied to recon-
figure switching points for average temperature calculation. The physical element
position will never be changed.

VH88 Short Error

Item type: read and write

Default: -49.5

Range: -49.5 ~ 359.5

Error message displayed when any element has a short circuit. Method of display can be

configured at VH92 "Error Display Select."

VH89 Open Error

Item type: read write

Default: 359.9

Range: -49.5 ~ 359.5

Error element displayed when any element has an open circuit. Method of display can be

configured at VH92 "Error Display Select."

8.1.9

Device setting 2: VH90~VH99

VH90 Device ID Number

Item type: read and write

Default: 0

Range: 0 ~ 16777214

In order to distinguish own device ID when 4532 ATC is connected in multi drop HART

loop.

Caution!

Changing device ID may lead to communication error because of mismatched pre-
registered device ID and HART address.

VH91 Previous Error

Item type: read only

Display of error history. Coded error message will be the same contents as VH80.