12 technical data – Varec 4532 User Manual

Page 61

Advertising
background image

Varec, Inc.

51

4532

Technical data

12 Technical data

Technical data at a glance:

Application

Application

The 4532 ATC performs precise liquid and gas phase average

temperature measurement in bulk storage tank applications.
• Standard 2" flange installation
• Overall 18.5m temperature measurement range

Function and system design

Measuring principle

Temperature measurement
• Up to 6 elements spaced evenly over the length of the flexible

tube.

• Pt100, Class A elements

Input

Measured variable

Temperature measurement
Temperature conversion range: -20 ~ +100°C

Measuring range

-20 ~ +100°C (-4 ~ 212°F)

Output

Output signal

• HART protocol (multi drop HART loop connection)

Signal on alarm

Error information can be accessed via the following interfaces and

transmitted digital protocol (refer to the operation manuals on

following instruments)
• Tank Side monitor 4590 TSM
• 6000 Series Servo Tank Gauge

Auxiliary energy

Load HART

Minimum load for HART communication: 250

W

Cable entry

see technical document

Supply voltage

16 ~ 30VDC (on multi drop HART loop)

Current consumption

Less than 6mA

Performance characteristics

Reference operating

conditions

• temperature = +25 °C (77 °F) ±5 °C (9 °F)
• pressure = 1013 mbar abs. (14.7 psia) ±20 mbar (0.3 psi)
• relative humidity (air) = 65 % ± 20%

Maximum measured

error

Typical statements for reference conditions, include linearity,

repeatability, and hysteresis:
• Linearity:

• Temperature: ±0.15°C (0.27°F) + number of element deviation

(based on IEC class A standard)