VIEW OR SET THE DYNAMIC CALIBRATION:

This mode allows you to set a resistance compensation factor

determined by the type of interface device used. When using a

nasal mask or a full face mask, press the up or down arrows

to select “r nr”. If using a nasal device such as the Nasal-Aire I or

II or the Spiritus, press the up or down arrows to select

“r r”, the blower function will compensate for device restriction.

Once the desired setting is reached, press the mode button to save

the setting.

For Tier 2, the display is now back to prescription pressure and

you may review all mode settings or turn the power off to exit

technician mode.

For Tier 3, the display will show “LOAd” only. This is for

factory use only. You may push mode to get back to prescription

pressure or turn the power off to exit technician mode.

WARNING:

The Tier 2 and 3 units are shipped with a dust cover on the DB9

connector. This cap should be left on except when the connector is

in use.

Tier 3 Only:

AUTO MASK CHANGE REMINDER INDICATOR & RESET:

The “TIME” Icon on the LCD flashes after 90 days total time.

To reset, press and hold the mode and down arrow button for

3 seconds. The “Time” Icon goes off and the 90 day time interval

re-starts. This feature can only be disabled with an external

command via the external software and DB9 connector.

Tier 1, 2 & 3:

To exit out of the technician mode on all Tier models, press the on/off
button at any time. The light will turn off.

10 RemRest