Activeforever RemRest 901 CPAP Machine User Manual

Page 8

Advertising
background image

TO VIEW OR RESET THE USAGE TIME:

The USE and TIME icons will flash on the LCD and the

usage time the device was in use by the patient will be displayed on

the LCD. This value should be zeroed for new patients to remove

factory test hours.
Press the up and down arrows at the same time to reset the

usage time. Once the desired setting is reached, press the mode

button to save the setting.

TO SET THE RAMP STARTING PRESSURE:

The default setting for the starting ramp pressure is 3.5 cmH2O.

The

LO icon will flash and the starting ramp pressure will be

displayed on the LCD. Press the up arrow or down arrow

to adjust the pressure to the desired start ramp pressure. The

pressure will adjust in increments of .1cmH2O. To move through the

settings more quickly, hold down the up or down arrow

button. Once the desired setting is reached, press the mode button

to save the setting.

TO SET THE RAMP TIME:

The

RAMP and

TIME icons will flash on the LCD and the

time in minutes will be displayed on the LCD. Press the up arrow

or down arrow to adjust the ramp time. The time will adjust in

5 minute increments from 5 to 50 minutes. Once the desired setting

is reached, press the mode button to save the setting.

TO SET THE AUTO COMPLIANCE:

The code “AC0” is displayed on the LCD indicating the factory

default setting of zero weeks. Press the up arrow or down

arrow to adjust the number of weeks from 1 to 26. Adjusting the

display to read AC0 or zero weeks turns off the auto compliance

feature. Once the desired setting is reached, press the mode button

to save the setting. The unit will perform a weekly check to confirm

28 hours of usage time per week (4 hours/day) for one to 26

weeks. If the client has failed to use the device, then the green

8

RemRest