Activeforever RemRest 901 CPAP Machine User Manual

Page 7

Advertising
background image

RemRest 7

To enter the Technician Mode, plug one end of the AC power cord

into the AC inlet plug located on the back of the RemRest device.

Plug the other end into the wall socket. The light will turn on for two

seconds then will turn off. The LCD will also turn on completely

(as shown above) for 2 seconds and will then display prescription

pressure.

With the motor blower off,

• hold down the “up” arrow and the “down” arrow

• simultaneously push and hold on/off button for two seconds

• the light should start blinking 1/2 second on and 1/2 second off

NOTE: First time device is put into service with a new patient, run

time, usage time and total time must be reset to insure accurate

compliance information.

NOTE: To advance through the different modes, press mode button

repeatedly until in the desired mode.

TO SET THE PRESCRIPTION PRESSURE:

The default setting is 10cmH2O. Press the up arrow or down

arrow to adjust the pressure to the prescribed setting. The

HI

will flash on the LCD and the pressure will be displayed cmH2O as

xx.x. The pressure will adjust in increments of 0.1cmH2O. To move

through the settings more quickly, hold down the up or down

arrow button. Once the desired setting is reached, press the mode

button to save the setting and enter next mode.

TO VIEW OR RESET THE RUN TIME:

The RUN and TIME icons will flash on the LCD and the

run time will be displayed on the LCD. Run time is the number of

hours the device is powered on. This value should be zeroed for new

patients to remove factory test hours.

Press the up and down arrows at the same time to reset the

run time to 0. Once the desired setting is reached, press the mode

button to save the setting.