Remrest set up, Tier 1 – Activeforever RemRest 901 CPAP Machine User Manual

Page 3

Advertising
background image

RemRest Set Up

Tier 1:

To enter the Technician Mode, plug one end of the AC power cord

into the AC inlet plug located on the back of the RemRest device.

Plug the other end into the wall socket. The green light will flash on

for two seconds then will remain off.

With the motor blower off,

• hold down the “up” arrow and the ”down” arrow on

(the Ramp button)

• simultaneously push and hold the on/off button for two seconds

• the on/off light should start blinking 1/2 second on and

1/2 second off for 5 seconds, indicating you are in technician

mode. It will then go right into Set Prescription Pressure Mode

(blink once then off 3 seconds).

ON/OFF

LIGHT

RAMP
LIGHT

RemRest 3

Advertising