Activeforever Invacare XPO2 Portable Oxygen Concentrator User Manual

Page 15

Advertising
background image

SECTION 5—TYPICAL PRODUCT PARAMETERS

Part No 1148112

15

Invacare XPO

2

Oxygen Concentration:*
*Based on an atmospheric
pressure of 14.7 psi (101 kPa) at
70°F (21°C)

87% to 95.6%, after initial warm-up
period (approximately 5 minutes)

Oxygen Flow:

Pulse flow delivery. Bolus Volume Ranges
From 300 - 840cc settings 1 to 5.

Dimensions:

10” high x 7” wide x 4” deep (25.4 cm
high x 17.8 cm wide x 10.2 cm wide)

Weight:

6.0 lbs (2.72 kg); 1.3 lbs (0.68 kg) for
supplemental battery

Battery Duration (times are
approximate):

Setting 1 = 3.5 hrs

Setting 2 = 2.5 hrs

Setting 3 = 2.0 hrs

Setting 4 = 1.5 hrs

Setting 5 = 1.0 hrs

Battery Recharge Time:
NOTE: Recharge time increases if 
battery is charging while unit is 
running.

4 hours

Humidity

Operating Humidity: 15% to 60% non
condensing

Storage Humidity: up to 95% non
condensing

Temperature Range:
(All power sources)

Operating temperature: 41°F to 95°F
(5°C to 35°C)

Storage temperature: -2°F to 140°F
(-20°C to 60°C)

Extended Temperature Range:
(Using AC or DC adapters)

Operating Temperature:
95°F to 104° (35°C to 40°C)
Continuous use - all settings

Extended Temperature Range
(Using Internal Battery)

Operating Temperature:
95°F to 104° (35°C to 40°C)
Continuous use - settings 1, 2 and 3
45 minutes (max) - setting 4
30 minutes (max) - setting 5