SECTION 6—OPERATING INSTRUCTIONS

Invacare XPO

2

34

Part No 1148112

System Too Hot/Cold Running Alarm – If the internal sensors 
register temperatures outside factory set levels during 
operation, the unit will alarm with a rapid audible beep, the 
red alarm indicator light will illuminate continuously, the unit 
will not operate and the flow setting 1 & 3 blue indicator 
lights will illuminate. The fan will turn On. Allow the unit to 
cool or warm to the recommended operating temperature 
range (see Recommendations for Optimal Performance on 
page 18).
 Turn Off the unit and try again. Change to an 
alternate source of oxygen while waiting. If the alarm 
continues, contact your equipment provider.

Battery Too Hot/Cold Alarm – If the internal battery sensor 
registers temperature outside a factory defined temperature 
range while the unit is operating, the unit will alarm with a 
rapid audible beep, the red alarm indicator light will 
illuminate continuously, the unit will stop running and the 
flow setting 1 & 4 blue indicator lights will illuminate. Turn 
Off the unit and disconnect AC or DC power adapters. Allow 
to the unit to cool or warm, to the recommended operating 
temperature range (see Recommendations for Optimal 
Performance on page 18)
 and try again. Change to an 
alternate source of oxygen while waiting. If the alarm 
continues, contact your equipment provider.

Stuck Button Alarm – During power up of the unit, if it detects 
that a button is stuck or being pressed too early, the unit will 
alarm with a rapid audible beep, the red alarm indicator light 
will illuminate continuously, the unit will not operate and the 
flow setting 1 & 5 blue indicator lights will illuminate. Turn 
Off the unit. Individually press all the front panel buttons, 
looking for buttons that seem stuck in a “pressed” position. 
Try turning On the unit again without pressing any other 
buttons during the power up. If the alarm continues, change 
to an alternate source of oxygen and contact your equipment 
provider.