SECTION 6—OPERATING INSTRUCTIONS

Part No 1148112

33

Invacare XPO

2

Battery Discharged Alarm – Once the XPO

2

’s internal battery 

(or supplemental battery, if applicable) capacity drops to 
minimum, a triple audible beep will sound and the GREEN 
25% battery gauge indicator will blink very rapidly. The unit 
will then shutdown.

No Breath Detect Alarm ‐ When the XPO

2

 is operating but 

does not sense breathing, a constant audible alarm sounds 
and the red alarm indicator light will illuminate continuously. 
If this occurs, check the connection from the cannula to the 
XPO

2

 unit and ensure that you are breathing through your 

nose. If the alarm continues, change to an alternate source of 
oxygen and contact your equipment provider. 

Breath Rate Over Capacity Alarm – If your breathing rate 
causes the capacity of the XPO

2

 to be exceeded, an 

intermittent audible beep sounds and the red alarm indicator 
will illuminate intermittently. When this occurs, reduce your 
activity and/or change to another source of oxygen. 

If your breathing rate continues to increase the audible beeps 
and red indicator blinks will become more rapid.
NOTE: Oxygen will be supplied during this alarm.

System Too Hot/Cold for Start Alarm – If the internal sensors 
register temperatures outside factory set levels upon start‐up, 
the unit will alarm with a rapid audible beep, the red alarm 
indicator light will illuminate continuously, the unit will not 
operate and the flow setting 1 & 2 blue indicator lights will 
illuminate. The fan will turn On. Allow the unit to cool or 
warm to the recommended operating temperature range (see 
Recommendations for Optimal Performance on page 18). 
Turn Off the unit and try again. Change to an alternate source 
of oxygen while waiting. If the alarm continues, contact your 
equipment provider.