SECTION 8—TROUBLESHOOTING GUIDE

Part No 1148112

43

Invacare XPO

2

OPERATING ALARM:
Continuous
audible beep and
red alarm
indicator is
illuminated.

AND

Flow indicators
3 & 4 are
illuminated.

Abnormal operation has
been detected.

1. Turn Off the unit.
2. Turn On the unit.
3. If the alarm continues,
change to another
source of oxygen and
contact your equipment
provider

COMPRESSOR ALARM:
Continuous
audible beep and
red alarm
indicator is
illuminated.

AND

Flow indicators
4 & 5 are
illuminated.

Abnormal compressor
operation has been
detected.

1. Turn Off the unit.
2. Turn On the unit.
3. If the alarm continues,
change to another
source of oxygen and
contact your equipment
provider

SYSTEM ALARM:
Continuous
audible beep and
red alarm
indicator is
illuminated.

AND

Flow indicators
3, 4 & 5 are
illuminated.

Abnormal system operation
has been detected.

1. Turn Off the unit.
2. Turn On the unit.
3. If the alarm continues,
change to another
source of oxygen and
contact your equipment
provider

ALARM/WARNING:
SYMPTOM

PROBABLE CAUSE SOLUTION