6. BATTERY CHARGING AND CARE


Cleaning the battery

If the batteries are contaminated by water, battery acid, dust or other substances, they will discharge

quickly. The batteries supplied with the power chair are sealed and as such are maintenance free with no

risk of battery leakage. Please follow the steps below to clean the battery.

1. Turn the Image EC power switch to OFF.
2. Remove the seat.
3. Remove the shroud and unplug the terminal of the taillight and signal lights.
4. Use a clean cloth to wipe off the soiled area.
5. Remove the battery.
6. Wipe with a clean cloth. If the terminal is covered with white powder, wipe it clean using warm water.

7. Do not touch the battery posts.


Batteries replacement

1. Remove the seat. (See To remove the seat”)
2. Remove the shroud.
3. Disconnect the negative terminal (black) wires first, then the positive terminal (red) wires by

removing the four nuts and bolts from each of the battery terminals.

4. Remove the batteries.
5. Place new battery.
6. Connect the positive terminal (red) wires first, then the negative terminal (black) wires by tightening

the nuts and each battery terminal.

7. Install the shroud and seat.

8. Do not touch the battery posts.


Warning

Be sure the battery wires are connected to the right battery terminal.

Suggestions

1. Make sure the terminals are installed properly and replace the cover.

2. Battery efficiency will vary with outside conditions- driving range will be reduced in cold

weather If the power chair is not used for a long time, charge the battery at least once a week.

3. Replace both batteries together.