Troubleshooting

15

Incorrect Operation

If system positioning does not seem to be repeatable, the problem could be with
the machine tool or with the readout system.

Check the machine tool.

Check that the table is not locked. Check the gib

adjustments. Check spindle run-out.

Check linear encoders.

Check each encoder and reading head for proper

installation. Ensure that the mounting brackets are secure.

If the system seems to be displaying incorrect positions, check the following items.

Verify grinding wheel or tool.

Check the grinding wheel or tool for wear.

Verify linear error compensation.

Make sure that the factors used for linear

error compensation are correct.

Power-On Self Test

There are four tests performed when power is applied to the readout. You will not
notice these tests unless a problem is found, in which case a message will scroll
across the display.

Keypad—a key was detected stuck down. The message will display what row (x)
and column (y) the stuck key is in:

KEY STUCK – RxCy – PRESS CLEAR

Parameter memory—some of the system setup parameters are not valid. You
may be able to check and reset these settings, but it’s possible they will be lost
again. The readout should be serviced as soon as possible.

MEMORY FAILURE [1] – PRESS CLEAR

!

CAUTION

Some working settings are not valid. Proceed with caution.