Power Supply Front Panel Layout

1.4 Power Supply Unit Front Panel Layout

The power supply individual front panels contain the input ON/OFF

circuit breakers and the FAULT, MASTER / STANDALONE and SLAVE

indicators. Refer to Figure 1-2 and Table 1-3. for a description of front

panel controls and indicators.

Figure 1-2.

Model 4Q06125PS Front Panel

Table 1-3.

Power Supply Front Panel Controls and Indicators

Reference

(Figure Fig-

ure 1-2.)

Control or Indi-

cator

Function

1

POWER ON/OFF

circuit breaker

switch

Applies source power to unit

2

SLAVE indicator

Lights when set as SLAVE Model 4Q06125PS in the
4Q06250PS configuration.

3

MASTER indicator

Lights when configured as MASTER Model 4Q06125PS
in the 4Q06250PS configuration

4

FAULT indicator

Lights red when a fault is detected. The following failure
or fault conditions can cause the FAULT indicator to light:
overtemperature, instant internal overcurrent, output
overvoltage/overcurrent, local 15V failure, input under/
overvoltage, input overcurrent, internal output under/
overvoltage, internal output overcurrent, overtempera-
ture, fan failure.
When the FAULT indicator lights, an audible beep
sounds a warning for approximately two seconds and the
output is crowbarred by an internal contactor. The fault is
latched. After the cause of the fault is removed, the unit
can be restarted by cycling the POWER circuit breaker to
OFF, then ON, or by applying a START_EXT pulse at
Analog I/O Port pin 7

SLAVE

MASTER / STANDALONE

FAULT