34

Rev. 5

Operation

Shift Key Commands : Ramp Rate

is installed), attempting to use the fine adjust knob for PSw P/S

ramp rate will produce one beep

1

.

Pressing

SHIFT

and then

RAMP RATE

will access the ramp rate

menu. The numeric and

ENTER

keys (or the fine adjust knob) are

used to set the segment 1 ramp rate to a value of 0.2.

The right arrow key is pressed once to access the segment 1 range

display. The numeric and

ENTER

keys (or fine adjust knob) are

used to set the segment 1 current range upper bound to a value of

55.

Pressing the right arrow key accesses the next (second) segment

ramp-rate display. The segment 2 ramp rate is set to a value of 0.1.

The right arrow key is pressed once to access the segment 2 range

display. The segment 2 current range upper bound is set to a value

of 58.

5. Since the PSw P/S ramp rate is active in that scenario, and not the segmented

ramp rate.

1. Since the standard segmented ramp rate is active in that scenario, and not the PSw

P/S ramp rate.

+50.00 A - Seg.1 Ramp Rate (A/sec)

+0.50 Vs

±0.2000

+50.00 A - Seg.1 Range (A)

+0.50 Vs

0.0 to ±55.0

+50.00 A - Seg.2 Ramp Rate (A/sec)

+0.50 Vs

±0.1000

+50.00 A - Seg.2 Range (A)

+0.50 Vs

±55.0 to ±58.0