5 type plate – VEGA VEGASWING 83A EXD User Manual

Page 14

Advertising
background image

14

VEGASWING 81A EXD and 83A EXD

Product description

1.5 Type plate

Before mounting and electrical connection,
please check if the correct instrument is
used. Note the type plate which is located as
follows:

1 Master data of the order no.
2 Ex-certification number/Ex-classification

Explosion protected version - note the
information and conditions in the certificate

3 Data of the electronics
4 No. of the order confirmation/Pos.-no.
5 Serial number
6 Test mark when used as par t of an over fill

protection for vessels storing water endan-
gering liquids - note the information and
conditions of the general type approval

7 Manufacturing year
8 Test authority no. CE

Order code

Detailed information to the order code is
stated in the "Product Information Vibration “
or in the "VEGA-pricelist “.

Type plate

The type plate contains important data re-
quired for mounting and connection. The
configuration and the parts of the type plate
are explained in the following example.

Configuration of the type plate (example)

type: S81GNN.RAK
see certificate
PTB 99 ATEX XXXX
II 2G EEx d IIC T6
electronics: R
power supply:

20…250VAC.
20…60VDC

protection: IP67

ser.no. XXXXXXXX

Ord.no.:123456/000

Insp.

1999

0032

1

2

8

3

4

5

7

6

Z-11.65-