5 diagnosis, 1 recurring test acc. to whg, 2 exchange of electronics – VEGA VEGASWING 83A EXD User Manual

Page 25

Advertising
background image

VEGASWING 81A EXD and 83A EXD

25

5.2 Exchange of electronics

• Interrupt the voltage supply and discon-

nect the connection lines and the switching
current circuit on the switching cabinet.

• Make sure that there is no Ex-atmosphere.
• Loosen the four screws of the housing

cover with a screwdriver and remove the
cover.

• Disconnect the connection lines on the

oscillator.

• Remove the old oscillator and insert the

new one.

• Set the A/B-switch to the same position as

on the old oscillator (not with SWING
E82 Z).

• Set the potentiometer for the density ad-

justment of the new electronics to the same
value as on the old oscillator.

• Fasten the housing cover and remove the

screw on the electronics box. Note that the
screw head is turned in one of the open-
ings.

• Switch on the power supply . The sensor is

immediately ready for operation.

Test course

A-mode

B-mode

1 Simulation of a fault signal (after release of key approx. 3 s)

1)

Level relay

deenergized > 22 mA

2)

Failure relay

deenergized < 2,2 mA

2)

2 Simulation of an empty signal (approx. 1,5 s)

Level relay

energized

8 mA

2)

Failure relay

energized

3 Simulation of a full signal (approx. 1,5 s)

Level relay

deenergized 16 mA

2)

Failure relay

energized

4 Return to the actual operating condition (covered/uncovered)

1)

as long as the key is pushed, the instrument signals failure.

2)

condition of VEGASWING

With the provided current values, you can also carry out the function test directly via a DCS or
a processing system

Diagnosis

• Connect the connection lines again.