3 measurement of liquids, Flat vessel top – VEGA VEGASON 54K…56K User Manual

Page 26

Advertising
background image

26

VEGASON 54K … 56K

< 100 mm

It is also possible to mount the sensors in
version C in a 1“ thread.

4.3 Measurement of liquids

Flat vessel top

On flat vessels, mounting is usually done on
a very short DIN socket piece. Reference
plane is the lower edge of the flange. The
transducer should protrude out of the flange
tube.

VEGASON 55 in flange version on short
DIN socket piece

Reference plane

Mounting and installation

Min. distance
Type 54: 1 m
Type 55: 0.8 m

< 400 mm

VEGASON 56 in flange version on short
DIN socket piece

Reference plane

Min. distance
1.8 m

A mounting location directly on the vessel top
is ideal. A round opening in the vessel top is
sufficient to fasten the sensor with the flange,
or version B and C with swivelling holder.

Flange version and swivelling holder on flat vessel
top

Reference
plane

Reference
plane

< 60 mm

Mounting of the transducer in 1“ thread

Reference
plane

Min. meas. distance

1.8 m

Min. meas. dis-

tance

1.4 m

Swivelling
holder

Advertising