2 application features, Continuous and accurate – VEGA VEGAPULS 51K…54K 4 … 20 mA; HART compact sensor User Manual

Page 5

Advertising
background image

VEGAPULS 51K … 54K

5

21750-EN-031222

Reflected radar power dependent on the dielectric
constant of the measured product

2

0

0

10

25 %

40 %

5

10

5 %

20

30

40

50

%

4

6

8

12

14

16

18

20

ε

r

Product description

All products which are electrically conductive
reflect radar signals very well. Even slightly
conductive products provide a sufficiently
strong reflection for a reliable measurement.

All products with a dielectric constant

ε

r

greater than 2.0 reflect radar pulses suffi-
ciently (note: air has a dielectric constant

ε

r

of

1). Signal reflectivity grows stronger with
increasing conductivity or increasing dielec-
tric constant of the product. Hence, nearly all
substances can be measured.

Temperature influence: Temperature error absolutely
zero (e.g. at 500°C 0.018 %)

100

500

1000

1300 ˚C

0

0

0,01

0,02

0,03

%

0,018 %

0,023 %

Pressure influence: Error with pressure increase very
low (e.g. at 50 bar 1.44 %)

10

0

0

5

%

1,44 %

2,8 %

10

50

20

30

40

60

100

70

80

90

110

120

130

140

bar

0,29 %

3,89 %

With standard flanges of DN 50 to DN 250,
ANSI 2“ up to ANSI 10“ or G 1½ A and 1½“
NPT, the sensor antenna systems can be
adapted to various products and measuring
environments.

The high-quality materials can also withstand
extreme chemical and physical conditions.
The sensors deliver stable, reproducible
analogue or digital level signals with reliability
and precision and have a long useful life.

Continuous and accurate

Unaffected by temperature, pressure and
atmosphere content, VEGAPULS radar sen-
sors measure quickly, accurately and con-
tinuously the levels of widely varying
products.

VEGAPULS 50 sensors allow radar level
measurement in plants where it was hitherto
unthinkable because of high costs.

1.2 Application features

Applications

• level measurement of any liquid, limited in

solids.

• measurement also in vacuum
• all slightly conductive materials and all

substances with a dielectric constant > 2.0
can be measured

• measuring range 0 … 20 m.

Two-wire technology

• power supply and output signal on one

two-wire cable (Loop powered)

• 4 … 20 mA output signal or HART

®

output

signal.

Rugged and abrasionproof

• non-contact
• high-resistance materials

Exact and reliable

• accuracy 1 mm.
• unaffected by noise, vapours, dusts, gas

compositions and inert gas stratification

• unaffected by varying density and tem-

perature of the medium

• measurement in pressures up to 40 bar

and product temperatures up to 200 °C.

Advertising