7 technical data, 1 technical data – VEGA VEGAPULS 51K…54K 4 … 20 mA; HART compact sensor User Manual

Page 51

Advertising
background image

VEGAPULS 51K … 54K

51

21750-EN-031222

1000

900

800

700

670

600

500

400

300

250

200

100

0

14

20

25

30

35

15

29

36

V

975

720

25,5

19,5

Adjustment
resistance
(HART

®

and

VEGACONNECT)

max. voltage limit

Ex ia

sensors

max. voltage limit

non-Ex

and

Ex d ia

sensors

min. voltage limit when using the HART

®

adjustment resistance:
- non-Ex and Ex ia sensors
- Ex d ia sensors

N

on-E

x and

E

x ia

Exd

ia

7 Technical data

7.1 Technical data

Power supply

Power supply

- four-wire sensor

24 V DC (20 … 72 V DC)

(non-Ex and Ex d ia)

230 V AC (20 … 253 V AC), 50/60 Hz
fuse 0.315 A TR

- two-wire sensor

24 V DC (14 … 36 V DC)

- two-wire Ex ia sensor

24 V DC (14 … 29 V DC)

- two-wire Exd ia sensor

24 V DC (20 … 36 V DC)

Permissible residual ripple of the power
supply with two-wire sensors

- 0 … 125 Hz

1 V

ss

- 125 Hz … 500 Hz

1.0 … 0.01 V

ss

declining linearly

- 500 Hz … 10 kHz

0.01 V

ss

Power consumption

- four-wire sensor

max. 130 mA

- two-wire sensor

max. 22.5 mA

Power consumption

- four-wire sensor

max. 2.1 W; 7.5 VA

- two-wire sensor

55 … 810 mW

Load

- four-wire sensor

max. 500 Ohm

- two-wire sensor

see diagram

Technical data

max. load

non-Ex

max. load

Ex d ia

max. load

Ex ia

Advertising