4 Setup

Electrical connection and setup can be sim-
plified by temporarily mounting the cover or
the indicating module on VEGABAR laterally
or downwardly displaced.

OPERATE

t

SPAN

ZERO

i

Key +: Change
value (increase

Key –: Change
value (de-
crease)

Setup

Indicating module

VEGABAR

8

7

6

5

4

3

2

1

4 ¼ 20 mA

DISPLAY

VEGADIS 10

12 ¼ 36 V DC E12

+ -

+

-

OK

Tighten screws lightly

4.1 Indicating module

Digital value
- 4 digits and decimal point
- basic function module: indicates a perma-

nent bar measurement range

- menu-driven module: allows individual

scaling of bar measurement range

The connection is made via a plug connection
acc. to the scheme in the "Electrical connec-
tion“ section.

4.2 Setup with module "Adjustment

of the basic functions“

Adjustment elements

Tendency
indication

Bar graph

Digital value

Adjustment system

• Select the required function with the rotary

switch.

• Change the value with the "+“ and "–“ keys.
• Reset rotary switch to OPERATE, the ad-

justed values are transferred to the
EEPROM memory. They remain there even
in case of voltage drop.

Adjustment

For adjustment of zero and span, an amme-
ter must be connected to terminals 1 and 3.
The measured value is identical to the output
current.

1 Adjustment of zero

(e.g. process pressure zero or vessel
empty)
• Set rotary switch to zero.
• By pushing the "+“ and "–“ keys simultane-

ously, the current jumps directly to 4 mA or
set it to 4 mA by pushing the "+“ and "–“
keys.

Adjustment range of zero:
-20 % … +95 % of nominal measuring range
(corresponds to turn up of up to +95 %)

Rotary switch: Select requested
function