VEGABAR 44 (HART

®

)

Instrument modification

6 Instrument modification

6.1 Exchange of adjustment mod-

ules

The modular construction of VEGABAR allows
retrofitting, interchange or removal of the
adjustment and display modules. Data previ-
ously saved (e.g. adjustment values) are not
stored in the adjustment module but in an
EEPROM of the electronics module and
therefore readjustment is not necessary. The
connection of the modules is made via a 4-
pole plug connection.

Exchange of the adjustment module

Removal of the adjustment module

Separate VEGABAR from the power
supply

Loosen the screws on the upper side of
the housing and remove the cover or the
display module

Loosen connection cables from the termi-
nals, if necessary pull out the plug con-
nection of the display module

Loosen the two screws of the adjustment
module

Remove adjustment module and pull out
the plug connection

Insertion of an adjustment module

Plug the plug connection of the new ad-
justment module into the plug-in socket of
the sensor electronics

Fasten the new adjustment module

Reconnect the connection cables, if nec-
essary, also the cable from the display
module

Close cover or display module of VEGA-
BAR

Reconnect VEGABAR to power supply

VEGABAR

Module
without
adjustment

Module for
menu-driven
adjustment

Module for
basic
functions

Display
module