5.8 Switch on phase
A
fter connecting VEGABAR 61 to power supply or after a voltage
recurrence, the instrument carries out a self-check for approx. 30
seconds:
l
I
nternal check of the electronics
l
I
ndication of the instrument type, the firmware as well as the
sensor TAGs (sensor designation)
l
O
utput signal jumps briefly (approx. 10 seconds) to the set fault
current
T
hen the corresponding current is outputted to the cable (the value
corresponds to the actual level as well as the settings already carried
out, e.g. factory setting).