4 adjustment with hart® handheld, 4 adjustment with hart, Handheld – VEGA VEGAPULS 43 4 … 20 mA; HART compact sensor User Manual

Page 48

Advertising
background image

48

VEGAPULS 43 – 4 … 20 mA

26626-EN-041227

Connection to a VEGA signal conditioning
instrument

If you operate a HART

®

compatible sensor on

a VEGA signal conditioning instrument, you
have to connect the sensor via a resistor
(see following table) during HART

®

adjust-

ment. This resistor, together with the internal
resistance of the instruments, provides the
value of 250 Ohm required for the HART

®

instrument. An inherent system load resist-
ance allows a corresponding reduction of Rx.

VEGA signal conditioning instr.

Rx

VEGAMET 513, 514, 515, 602

50 … 100 Ohm

VEGAMET 614

no additional

VEGADIS 371

resistor
required

VEGAMET 601

200 … 250 Ohm

VEGASEL 643

150 … 200 Ohm

VEGAMET 513 S4, 514 S4
515 S4, VEGALOG EA card

100 … 150 Ohm

Set-up

Just connect the HART

®

handheld to the

signal cable, after having connected the
sensor to power supply.

Note:

If the resistance of the power supply is less
than 250 Ohm, a resistor must be connected
into the signal/connection loop during adjust-
ment.

The digital adjustment and communication
signals would otherwise be short-circuited
due to insufficient resistance of the supply
current source or the processing system,
and as a result, communication with the sen-
sor would not be ensured.

+

-

Ri

≥ 250 Ω

250

+

-

Ri

<

250

With any HART

®

handheld you can set up the

VEGAPULS series 40K radar sensors like all
other HART

®

compatible sensors. A special

DDD (Data Device Description) is not neces-
sary.

5.4 Adjustment with HART

®

handheld

Advertising