54

VEGAPULS 43 – 4 … 20 mA

26626-EN-041227

Technical data

Characteristics

1)

(typical values under reference conditions, all statements relate to the nominal measuring
range)

Min. span between

full and empty

> 10 mm (recommended > 50 mm)

Meas. frequency

26 GHz technology

Intervals

- two-wire sensor (4 … 20 mA)

1 s

- two-wire sensor (digital)

0.6 s

- four-wire sensor

0.5 s

Beam angle (at -3 dB)

- DN 50, ANSI 2“

18°

- DN 80, ANSI 3“

10°

- DN 100, ANSI 4“

10°

- DN 150, ANSI 6“

10°

- TRI-Clamp 2“

18°

- TRI-Clamp 3“

18°

- socket DN 50

18°

- socket DN 80

18°

Adjustment time (response time)

2)

> 1 s (dependent on the parameter adjustment)

Influence of the process temperature

cannot be measured at 0 bar;
at 5 bar 0.004 %/10 °K;
at 40 bar 0.03 %/10 °K

Influence of the process pressure

0.0265 %/bar

Adjustment time

2)

> 1 s (dependent on the parameter adjustment)

Emitted radar power (average)

0.717 µW

Received average emitted power

3)

- distance 1 m

0.4 … 3.2 nW per cm² (0.4 … 3.2 x 10

-9

W/cm²)

- distance 5 m

0.02 … 0.13 nW per cm²

Ex technical data

Comprehensive data in the safety instructions manual (yellow binder)

WHG approvals

VEGAPULS 43 radar sensors are approved as part of an overfill protection system for
stationary vessels storing water-endangering liquids.

Materials

Housing

PBT (Valox) or
Aluminium die casting (GD-AlSi 10 Mg)

Connection housing with Exd version

Aluminium mould casting (GK-AlSi 7 Mg)

Process flange

1.4435

Antenna (wetted parts)

TFM-PTFE

1)

Similar to DIN 16 086, reference conditions acc. to IEC 770, e.g.
temperature 15 °C … 35 °C; moisture 45 % … 75 %; pressure 860 mbar … 1060 mbar

2)

The adjustment time (also actuating time, response time or adjustment period) is the time required by the
sensor to output the correct level (with max. 10% deviation) after a sudden level change.

3)

Average emitted power (electromagnetic energy) received by a body per cm² directly in front of the antenna.
The received emitted power depends on the antenna version and the distance.