VEGA D96_D97 Profibus PA User Manual

Page 26

Advertising
background image

26

Pressure transmitter D96/D97 (Profibus PA)

If you want to carry out the "Adjustment tak-
ing the actual product into account" (live
adjustment), click in this window to

yes

, then

to

OK

.

Setup

Here you can choose in which unit the adjust-
ment should be made (mbar or bar, psi, kPa,
inHg). Then you enter the pressure values
corresponding to 0 % and 100 %. Then click
to

OK

. The adjustment is now finished.

Click to

Min-Adjustment

or to

Max-Adjust-

ment

to carry out the respective adjustment.

Note:

For min. adjustment and max. adjustment,
defined levels (e.g. 0 % and 100 % or e.g.
20 % and 80 %) have to be present in the
vessel.

In the window "Min-Adjustment" you choose in
which unit the adjustment should be carried
out (bar, mbar, psi, kPa, inHg). Then click to

Save

. The min. adjustment is finished.

In the window "Max-Adjustment" you choose
in which unit the adjustment should be car-
ried out (bar, mbar, psi, kPa, inHg). Then
click to

Save

. The max. adjustment is fin-

ished.

Sensor parameter setting – Linearisa-
tion curves

Click in the window "Instrument data param-
eter adjustment" to

Conditioning

.

Advertising