4 Mounting

4.1 General instructions

If the electronics module is defective, it can be replaced by the

user.

In Ex applications, only one instrument and one oscillator with

appropriate Ex approval must be used.

If there is no electronics module available on site, it can be

ordered from the responsible VEGA agency.

The order data and company data of the sensor must be

loaded in the new oscillator. These are the options:

l

at the factory by VEGA personnel

l

or on site by the user.

Information:

When loading on site, first of all the order data must be

downloaded from the Internet (see "Setup").

In both cases, the sensor serial number is necessary. The

serial numbers are stated on the type label of the instrument,

inside the housing or on the delivery note.

Caution:

The order data and company data contain important preset-

tings for the sensor. A reliable operation and a correction of the

measurement is not possible without these data.

Electronics modules are adapted to the respective sensor.

They differ, e.g. in the signal output, the power supply or

approval.
First of all check by means of the overview in section Mounting

preparations, if you are using the suitable electronics module.

Compare the new electronics module with the existing one.

The specifications on the type label must correspond. This

applies mainly to instruments with approvals.

4.2 Mounting preparations VEGAFLEX 61, 62, 65,

66

Electronics module FX-E.60H. suitable for VEGAFLEX 61, 62,

65, 66 – 4 … 20mA/HART:

Sensor serial number

Assignment

4 … 20 mA/HART

6

Electronics module - VEGAFLEX series 60

Mounting