3 exchanging the electronics module, 4 software update – VEGA VEGAPULS 61 (≥ 2.0.0 - ≤ 3.8) Profibus PA User Manual

Page 51

Advertising
background image

?

E

041, E042, E043

l

Hardware error, electronics defective

à Exchange the instrument or send it in for repair

?

E

113

l

Communication conflict

à Exchange the instrument or send it in for repair

Depending on the failure reason and measures taken, the steps
described in chapter "Set up" must be carried out again, if necessary.

8.3 Exchanging the electronics module

If the electronics module is defective, it can be replaced by the user.

In Ex applications only one instrument and one electronics module with
respective Ex approval may be used.

If there is no electronics module available on site, one can be ordered
from the VEGA agency serving you.

The new electronics module must be loaded with the settings of the
sensor. These are the options:

l

At the factory by VEGA

l

Or on site by the user

In both cases, the sensor serial number is needed. The serial numbers
are stated on the type label of the instrument, inside the housing or on
the delivery note.

Information:
When loading on site, first of all the order data must be downloaded
from the Internet (see operating instructions manual "Oscillator").

The electronics modules are adapted to the respective sensor and
distinguish also in the signal output or power supply.

8.4 Software update

The software version of VEGAPULS 61 can be determined as follows:

l

on the type label of the electronics

l

via the indicating and adjustment module

l

via PACTware

You can view all software histories on our website www.vega.com.
Make use of this advantage and get registered for update information
via e-mail.

Reaction after fault rec-
tification

Sensor serial number

Assignment

VEGAPULS

61 • Profibus PA

51

8 Maintenance and fault rectification

28444

-EN