VEGA VEGAPULS 61 (≥ 2.0.0 - ≤ 3.8) Profibus PA User Manual

Page 60

Advertising
background image

1,0 m

20 m

30 mm

15 mm

-15 mm

-30 mm

Fig. 43: Accuracy VEGAPULS 61 with plastic horn antenna and increased sensitivity

Influence of the ambient temperature to the sensor electronics

12)

Average temperature coefficient of the zero
signal (temperature error)

0.03 %/10 K

Ambient conditions

Ambient, storage and transport temperature

-40 … +80 °C (-40 … +176 °F)

Process conditions

Vessel pressure

-

Encapsulated antenna system

-100 … 300 kPa/-1 … 3 bar (-14.5 … 43.5 psi)

-

Plastic horn antenna

-100 … 200 kPa/-1 … 2 bar (-14.5 … 29.0 psig)

Process temperature (measured on the
process fitting)

-40 … +80 °C (-40 … +176 °F)

Vibration resistance

13)

-

Encapsulated antenna system

mechanical vibrations up to 4 g in the frequency
range 5 … 200 Hz

-

Plastic horn antenna with compression
or adapter flange

mechanical vibrations up to 2 g in the frequency
range 5 … 200 Hz

-

Plastic horn antenna with mounting strap

mechanical vibrations up to 1 g in the frequency
range 5 … 200 Hz

Electromechanical data - version IP 66/IP 67 and IP 66/IP 68; 0.2 bar

Cable entry/plug

14)

-

Single chamber housing

l

1 x cable gland M20 x 1.5 (cable: ø 5 … 9 mm),
1 x blind stopper M20 x 1.5

12)

Relating to the nominal measuring range, in the temperature range
-40 … +80 °C .

13)

Tested according to the guidelines of German Lloyd, GL directive 2.

14)

Depending on the version M12 x 1, according to DIN 43650, Harting, 7/
8" FF.

60

VEGAPULS

61 • Profibus PA

10 Supplement

28444

-EN

-100426