0,5 m (1.6 ft)
5,85 m (19.19 ft)
10 mm (0.394 in)
3 mm (0.116 in)
- 3 mm (- 0.116 in)
- 10 mm (- 0.394 in)
0
Fig. 16: Deviation VEGAPULS 62 standpipe
Influence of the ambient temperature to the sensor electronics
7)
A
verage temperature coefficient of the zero
signal (temperature error)
0.03 %/10 K
Influence of the superimposed gas and pressure to the accuracy
T
he spreading speed of the radar impulses in gas or vapour above the product is reduced by high
pressures. This effect depends on the superimposed gas or vapour and increases with low
temperatures. The following chart shows the deviation caused by some typical gases or vapours.
T
he stated values refer to the distance. Positive values mean that the measured distance is too
high, negative values that the measured distance is too small.
Gas phase
Temperature
1 bar/14.5 psig
10 bar/145 psig
50 bar/725 psig
A
ir/Nitrogen
20 °C/68 °F
0.00 %
0.22 %
1.2 %
A
ir/Nitrogen
200 °C/392 °F
0.00 %
0.13 %
0.74 %
H
ydrogen
20 °C/68 °F
-0.01 %
0.10 %
0.61 %
H
ydrogen
200 °C/392 °F
-0.02 %
0.05 %
0.37 %
W
ater (saturated
steam)
100 °C/212 °F
0.20 %
-
-
W
ater (saturated
steam)
180 °C/356 °F
-
2.1 %
-
Ambient conditions
A
mbient, storage and transport temperature
-40 … +70 °C (-40 … +158 °F)
7)
R
elating to the nominal measuring range, in the temperature range
-40 … +80 °C .