VEGA VEGAPULS 62 (≥ 2.0.0 - ≤ 3.8) standpipe ver. 4 … 20 mA_HART four-wire User Manual

Page 44

Advertising
background image

Process conditions

P

rocess temperature (measured on the process fitting), depending on seal and antenna system

-

FKM (V

iton)

-40 … +200 °C (-40 … +392 °F)

-

FFKM (K

alrez 6375)

-20 … +200 °C (-4 … +392 °F)

-

FFKM (K

alrez 2035, 6230)

-15 … +200 °C (+5 … +392 °F)

FKM (V

iton) with impedance cone PP

-40 … +80 °C (-40 … +176 °F)

FFKM (K

alrez 6375) with impedance cone

PP

-15 … +80 °C (+5 … +176 °F)

F

or the vessel pressure, you also have to note the specifications on the type label. Always the

lowest value is applicable.

V

essel pressure

-100 … 4000

kPa/-1 … 40 bar (-14.5 … 580 psig)

V

essel pressure relating to the flange nom-

inal stage

see supplementary instructions manual "Flanges
according to DIN-EN-ASME-JIS
"

V

ibration resistance

mechanical vibrations with 1 g and 5 … 100 Hz

8)

Electromechanical data - version IP 66/IP 67

C

able entry

-

D

ouble chamber housing

l

1

x cable gland M20 x 1.5 (cable: ø 5 … 9 mm),

1

x blind stopper M20 x 1.5; plug M12 x 1 for

VEGADIS

61 (

optional)

or:

l

1

x closing cap ½ NPT, 1 x blind stopper

½

NPT,

plug M12 x 1 for VEGADIS 61

(

optional)

or:

l

1

x plug (depending on the version), 1 x blind

stopper M20 x 1.5; plug M12 x 1 for VEGADIS
61 (

optional)

S

pring-loaded terminals for wire cross-sec-

tion

2.5

mm² (AWG 14)

Indicating and adjustment module

V

oltage supply and data transmission

through the sensor

I

ndication

LC

display in dot matrix

A

djustment elements

4

keys

P

rotection

-

unassembled

IP

20

-

mounted into the sensor without cover

IP

40

8)

T

ested according to the regulations of German Lloyd, GL directive 2.

44

VEGAPULS

62

standpipe version • 4 … 20 mA/HART four-wire

10 S

upplement

28738

-EN

-090305

Advertising