Information:
When loading on site, first of all the order data must be downloaded
from the Internet (see operating instructions manual "Oscillator").

The electronics modules are adapted to the respective sensor and
distinguish also in the signal output or power supply.

8.4 Software update

The following components are required to update the sensor software:

l

Sensor

l

Power supply

l

VEGACONNECT

l

PC with PACTware

l

Current sensor software as file

At "www.vega.com/downloads" go to "Software". Select under "plics
instruments and sensors
" the suitable instrument series. Load the zip
file via the right mouse key with "Save target as" e.g. on the desktop of
your PC. Extract all files available in the zip file, e.g. to the desktop.

Connect the sensor to power supply and provide connection from PC
to the instrument via VEGACONNECT. Start PACTware and provide
connection to the sensor, e.g. via the VEGA project assistant. Close
the parameter window of the sensor, as far as open.

Go in the PACTware menu bar to "Instrument data", "Additional
functions
" and "Update instrument software".

PACTware checks now the actual hardware and software version of
the sensor and displays the data. This procedure lasts approx. 60 s.

Push the button " Update software" and select the previously
extracted hex file. Then the software update can be started. Further
files are installed automatically. Depending on the sensor, this
procedure lasts approximately 1 h.

8.5 Instrument repair

If a repair is necessary, please proceed as follows:

You can download a return form (23 KB) from our Internet homepage
www.vega.com under: "D

ownloads - Forms and certificates - Repair

form".

By doing this you help us carry out the repair quickly and without
having to call back for needed information.

l

Print and fill out one form per instrument

l

Clean the instrument and pack it damage-proof

l

Attach the completed form and, if need be, also a safety data
sheet outside on the packaging

Assignment

Load sensor software to
PC

Prepare update

Load software into sen-
sor