VEGA VEGAPULS 65 (≥ 2.0.0 - ≤ 3.8) Foundation Fieldbus User Manual
Page 50
Vessel pressure
-
Screwed part PVDF
-100 … 300 kPa/-1 … 3 bar (-14.5 … 43.5 psig)
-
Screwed part and flange 316L
-100 … 1600 kPa/-1 … 16 bar (-14.5 … 232 psig)
Process temperature (measured on the process fitting)
-
Process fitting, thread PVDF
-40 … +130 °C (-40 … +266 °F)
-
Other process fittings
-40 … +150 °C (-40 … +302 °F)
Vibration resistance
mechanical vibrations with 4 g and 5 … 100 Hz
7)
Electromechanical data - version IP 66/IP 67 and IP 66/IP 68; 0.2 bar
Cable entry/plug
8)
-
Single chamber housing
l
1 x cable gland M20 x 1.5 (cable: ø 5 … 9 mm),
1 x blind stopper M20 x 1.5
or:
l
1 x closing cap M20 x 1.5; 1 x blind stopper
M
20 x 1.5
or:
l
1 x closing cap ½ NPT, 1 x blind plug ½ NPT
or:
l
1 x plug (depending on the version), 1 x blind
stopper M20 x 1.5
-
Double chamber housing
l
1 x cable entry M20 x 1.5 (cable: ø 5 … 9 mm),
1 x blind stopper M20 x 1.5; 1 x blind stopper
M
16 x 1.5 or optionally available with 1 x plug
M
12 x 1 for VEGADIS 61
or:
l
1 x closing cap ½ NPT, 1 x blind stopper
½ NPT
, 1 x blind stopper M16 x 1.5 or
optionally 1 x plug M12 x 1 for VEGADIS 61
or:
l
1 x plug (depending on the version), 1 x blind
stopper M20 x 1.5; 1 x blind stopper M16 x 1.5
or optionally available with 1 x plug M12 x 1 for
VEGADIS
61
Spring-loaded terminals for wire cross-sec-
tion
> 2
.5 mm² (AWG 14)
7)
Tested according to the regulations of German Lloyd, GL directive 2.
8)
Depending on the version M12 x 1, according to DIN 43650, Harting, 7/
8" FF.
50
VEGAPULS
65 • Foundation Fieldbus
10 Supplement
28452
-EN
-090305