VEGA VEGASON 65 Profibus PA User Manual
Page 53
Electromechanical data
Cable entry
-
Double chamber housing
l
1 x cable gland M20 x 1.5 (cable: ø 5 … 9 mm),
1 x blind stopper M20 x 1.5; plug M12 x 1 for
VEGADIS
61 (optional)
or:
l
1 x closing cap ½ NPT, 1 x blind stopper
½ NPT
, plug M12 x 1 for VEGADIS 61
(optional)
or:
l
1 x plug (depending on the version), 1 x blind
stopper M20 x 1.5; plug M12 x 1 for VEGADIS
61 (optional)
Spring-loaded terminals for wire cross-sec-
tion up to
2.5 mm² (AWG 14)
Transducer cable
10)
-
Length
5 … 300 m (16.4 … 984.3 ft)
-
Diameter
7.2 … 7.6 mm (0.283 … 0.299 in)
Indicating and adjustment module
Voltage supply and data transmission
through the sensor
Indication
LC display in dot matrix
Adjustment elements
4 keys
Protection
-
unassembled
IP
20
-
mounted into the sensor without cover
IP
40
Materials
-
Housing
ABS
-
Inspection window
Polyester foil
Power supply
Operating voltage
-
Non-Ex and Ex-d instrument
20 … 72 V DC, 20 … 253 V AC, 50/60 Hz
Power consumption
max. 4 VA; 2.1 W
Electrical protective measures
Protection
IP
66/IP 67
Overvoltage category
III
Protection class
I
10)
With separate version C and D.
VEGASON
65 • Profibus PA
53
10 Supplement
28788
-
EN
-081127